R&S®R-Line Compact Test Chamber
Precise measurements of radiated emissions of wireless terminals 800 MHz to 18 GHz
Specifications in Brief
General data
| Environmental conditions | ||
| Operating temperature range | +0°C to +35°C | |
| Storage temperature range | –10°C to +45°C | |
| Relative humidity | without condensation | 90% at +25°C |
| Power supply | 208 V to 230 V/50 Hz to 60 Hz | max. 350 VA |
| Environmental test | ||
| Electromagnetic compatibility | in line with EN55011 class B, EN55022:1987, class B | |
| Safety class | in line with EN60204-1 | |
| Mechanical data | ||
| Dimensions (W × H × D) | including four movable and lockable castors | 1690 mm × 1560 mm × 2130 mm 66,535 in × 61,417 in × 83,858 in |
| Weight | 400 kg 881,85 lb | |
| Door size (W × H) | 500 mm × 1000 mm 19,685 in × 39,37 in | |
| Size of EUT (W × H × D) | max. 200 mm × 200 mm × 200 mm max. 7,874 in × 7,874 in × 7,874 in | |
| Weight of EUT | max. 2 kg max. 4,41 lb | |
| Positioning speed of turntable | azimuth: 4,5 rpm to 13 rpm elevation: 0,1 rpm to 0,5 rpm | |
| Distance between test antenna and center of EUT | 1,02 m 40,157 in | |
| Interfaces | Data | |
| IEC/IEEE bus | in line with IEC 625-2 (IEEE 488,2) | for control of positioner |
| Test antenna | 2 × SMA (female) | |
| Communications antenna | 1 × N (female) | |
| RF feedthroughs | 2 × N (female) 2 × SMA (female) | for calibration or connecting the EUT |
RF specifications
| Measurement range | ||
| Frequency range | 800 MHz to 18 GHz | |
| Polarization | horizontal and vertical, through cross-polarized antenna | |
| RF power | max. 3 W | |
| Communications antenna | 800 MHz to 6 GHz, circularly polarized | |
| Field uniformity | meets CISPR 16-1-4:2007 | site VSWR <4 dB |
| max. EUT volume for uniform field | 200 mm× 200 mm × 200 mm | |
| volume (W × H × D) | max. 7,874 in × 7,874 in × 7,874 in |